Analog instruments cannot directly interface with digital computers; they require ADC conversion for data acquisition systems.
E₂/E₁ = L₂/L₁; E₂ = (107.8/106.4) × 1.018 ≈ 1.031 V.
At null position in LVDT, the core is centered, and secondary voltages from both coils cancel out, resulting in zero or minimum output.
EMF at 25°C = 1.0186 - 0.0004 × (25-20) = 1.0186 - 0.002 = 1.0166 ≈ 1.0170 V.
Megohmmeters use high DC voltages (1-5 kV) to force sufficient current through high resistances for measurable deflection.
For AC bridges at balance condition: Z₁ × Z₄ = Z₂ × Z₃, where Z represents impedance of arms.
Digital frequency counters provide accuracy of ±1 count with real-time display and are industry standard for this frequency range.
Power = V × I × cos(φ) = 250 × 4 × 0.8 = 800 W. The wattmeter reading equals actual power consumed.
Shunt resistance Rs = (Ig × Rg)/(I - Ig) = (5×10^-3 × 10)/(5 - 5×10^-3) ≈ 0.01 Ω
Null deflection (or balance) method is preferred because it doesn't depend on meter calibration or linearity. The measurement relies on the balance condition formula only.